Skip to content
For the complete documentation index, see llms.txt. This page is also available as Markdown.

Find and close compliance gaps

MSI gives you two ways to find what's missing from your documentation. The Gap Register tracks every element of the FDA premarket cybersecurity gap assessment against your documents. AINN Analysis reads an FDA additional information letter and checks whether your updated documents answer each of the FDA's questions.

Before you begin

  • Select the device and device version you're working on, and upload its documents.

  • For AINN Analysis, have the FDA letter as a PDF or .txt file, and put your updated submission documents in a document collection.

  • The Gap Register isn't available for every organization. If you don't see Gap Register in the sidebar, contact Medcrypt.

Track gaps in the Gap Register

The Gap Register lists every requirement in the gap assessment, grouped by regulatory section, with an outcome for each based on what MSI found in your documents.

  1. In the sidebar, select Gap Register.

  2. If the register says it hasn't been assessed yet, select Regenerate, or run a regulatory audit to populate it. The register also updates on its own when, for example, you upload a document or complete an audit.

  3. Choose a tab:

    • Register: every gap element, organized by section.

    • Outstanding: only the gaps that still need attention, with what's Missing and the suggested Remediation for each.

  4. Narrow the list with Search gaps... and the Outcome, Review, and Assignee filters. Use Sort by to order rows by Element, Outcome, or Review State.

Outcome
Meaning

Fully Addressed

Your documents cover the element.

Partially Addressed

Some evidence was found, but not enough.

Missing

No evidence was found in the documents attached.

Not Assessed

The element hasn't been checked yet.

Not Applicable

The element doesn't apply to this device, with a recorded reason.

Close a gap

  1. Select View Details on a gap, or select the row. The details show the Requirement, the Current Evidence, the Recommendation, where the evidence is Expected In, and any Evidence Documents MSI found.

  2. To add evidence, select Upload Documents, choose a Collection, and upload the files. Keep Attach this collection to the device after upload selected so MSI reads the new documents into the device's knowledge base.

  3. To assign the gap to a teammate, use the assign control under Assigned To.

  4. If the outcome is wrong, select Override Outcome, choose the Outcome, and give a Reason. A reason is required for Not Applicable. To undo it, select Remove Override.

  5. When you've checked the gap, select Mark as Reviewed.

When an automated source suggests a new outcome for a gap, the details show a Gap Status Proposal. Select Accept Proposal or Reject Proposal.

To download the register, select Export.

Respond to an FDA additional information request

During a premarket review, the FDA can put the review on hold and send an additional information request, often called an AI letter or AINN letter. It lists specific deficiencies, or asks, that you must answer before the review continues. AINN Analysis breaks the letter into individual asks and checks each one against your updated documents.

  1. In the sidebar, select Deliverables, then select the AINN Analysis tab. The same tab also appears in the Gap Register.

  2. Under AINN Letter, upload the FDA letter (PDF or .txt).

  3. In New Document Collection, select the collection with your updated submission documents.

  4. Optional: in Old Document Collection (Optional), select the collection with your original submission, for comparison.

  5. Select Analyze. Progress moves through Parsing Letter, Fast Analysis, and Deep Analysis.

The tab shows a warning icon when an analysis has items for you to review.

Understand the results

The results table lists each FDA Ask with its Category, Severity, Status, Confidence, and Remediation. Each ask's status is Addressed, Partially Addressed, or Not Addressed. Expand a row to see the analysis, the citations from your documents it relied on, and the status history.

To see each cited excerpt and whether it was verified word for word against your documents, select Traceability.

Review and respond

  1. On an ask, select Update Status.

  2. Choose the Status, explain what evidence supports it in Justification, and select Submit Review.

  3. When you're ready to answer the FDA, select Export, then:

    • Export PDF or Export DOCX to download the analysis.

    • Generate FDA Response to draft a response, then Export Response Letter to download it once it's ready.

To analyze another letter, select New AINN Analysis. Earlier analyses are listed under Previous Analyses.

Next steps

Last updated

Was this helpful?